Open original FDA record ↗Recall record
- FDA identifier
- Z-1064-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Feb 20, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
In September 2011 firm issued two Customer Safety Advisory Notices for syngo Imaging XS. Firm became aware of an unintended behavior when using syngo Imaging XS, version VA70A and VA70B with Client/Server Installations. Loading additional images/series from the Patient Browser using the thumbnails may cause unintended deletion of images/series. In response, firm issued a Customer Safety Advisor
Source facts
- Action
- Siemens Medical issued an "UPDATE INSTRUCTIONS SY055/11/S/UPDATE PACKAGE VA70B_0211" to all affected customers. The package described the product, problem, and instructions for customers to follow.
- Initiated
- 2012-01-26
- Root cause
- Software design
- Terminated
- 2012-09-29
- Distribution
- Nationwide Distribution-including the states of CA, CO, CT, FL, GA, IL, IN, IA, KS, MD, MN, MO, NE, NV, NJ, NY, NC, OH, PA, TN, TX, and WI.
- Recall number
- Z-1064-2012
- Recall status
- Terminated
- CDRH record ID
- 107231
- FDA classified
- 2012-02-20
- Recall event number
- 61071