Open original FDA record ↗Recall record
- FDA identifier
- Z-1064-2019
- Company in FDA record
- Keystone Dental Inc
- FDA record creation date
- Mar 27, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
Source facts
- Action
- Keystone Dental initiated recall by phone to each one of the impacted customers. The phone call was then followed by a Recall Notification Letter that was mailed via overnight FedEx mail on February 22, 2019. Letters states problem and action to take: Replacement 1.25mm Hex Driver(s) have been provided to you with this letter. The 1.25mm Hex Driver(s) currently in your kit should be discarded. Please refer to the images on page 2 to locate the driver within your kit. This issue is limited to this 1.25mm Hex Driver(s) only and no other product within the kit is affected by this issue. Complete the Reply Communication form.
- Initiated
- 2019-02-22
- Root cause
- Process control
- Terminated
- 2020-06-19
- Distribution
- FL, MA, SC, VA
- Recall number
- Z-1064-2019
- Recall status
- Terminated
- CDRH record ID
- 171470
- FDA record created
- 2019-03-27
- Recall event number
- 82374