Reason stated in the source
The sunlamp products failed to incorporate a timer system and a control for termination of radiation emission in accordance with the Federal Sunlamp Product Performance Standard.
Source facts
- Action
- Soleil Systems sent recall notification letters dated 6/28/04 to their dealers on 8/2/04, informing them of the noncompliances of failing to incorporate a permanently mounted timer and on/off switch in the sunbeds. Soleil will bring each sunbed into compliance and requested the dealers to review their sales records for any Sollux beds received between 4/28/04 and 11/01/03, and supply Soleil with a list of names and addresses of end users who received those units. Soleil sent out end user notifications dated 8/20/04 beginning on 8/23/04 and initiated the retrofit of the sunbeds on 10/25/04.
- Initiated
- 2004-08-02
- Root cause
- Other
- Terminated
- 2006-03-28
- Distribution
- The sunbeds were distributed through dealers in Missouri, Texas, Virginia, Kentucky and Illinois
- Recall number
- Z-1065-04
- Recall status
- Terminated
- CDRH record ID
- 34451
- FDA record created
- 2015-06-25
- Recall event number
- 29737