Z-1065-2019: Accelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage:

Recall record Z-1065-2019 for Accelerate Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1065-2019
Company in FDA record
Accelerate Diagnostics Inc
FDA record creation date
Mar 27, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to mitigate any risk associated with these rare isolates, the firm has implemented an Expert Rule that suppresses the meropenem result for isolates that display a profile consistent with those that generated a false-susceptible result. The firm has responded to this risk by implementing a set of suppression rules, in a software update, to prevent incorrect results.

Source facts

Action
On 03/13/2017, customers were notified via phone of the potential risk of very major errors with meropenem for systems that have not been equipped with the most recent software update. TB000037, dated 03/16/17, was sent to customers after product launch. TB000082 and TB000083, dated 06/08/2018, were sent to customers that ran patient samples prior to installing the 1.3.1 PSW000012C patch. Starting on 01/25/19, Medical Device Correction notices were sent to customers, which provided a secondary notification of the suppression rules previously provided in TB000082 and TB000083. Customers were asked to do the following: (1) If you have not already done so, please contact customer support to schedule installation of the latest software version on your system. (2) Retrospectively review affected runs, specifically meropenem results. (3) As an interim measure, and until the latest software version is installed, you should review meropenem results obtained with the original software and apply the AST suppression rules (as noted in technical bulletins TB000082 and TB000083) to those results. Customers were asked to compete and return an acknowledgement form to indicate that they have received the correction notice and taken appropriate action. North America Customer Service and Technical Support: Toll free: +1 (888) 586 2939 opt. 4, Direct: +1 (520) 365 3100, Email: Support@axdx.com. EMEA Customer Service and Technical Support:: Direct: +34 93 220 4547, Email: Support.emea@axdx.com.
Initiated
2017-03-13
Root cause
Software design
Terminated
2021-05-21
Distribution
Worldwide Distribution - US Nationwide North America and Middle East US: TX, IL, TN, GA, CA, AR, MD, VA, Washington DC, IA, PA, NY, WA, and MO. OUS (Foreign): UK, Spain, Italy, and Germany
Recall number
Z-1065-2019
Recall status
Terminated
CDRH record ID
171153
FDA record created
2019-03-27
Recall event number
82216

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗