The coating of laser mirrors in a defined manufacturing lot of the Visulink 532/U devices may partially tear open and loosen from the base plate. The mirror may misdirect the laser beam to an unintended target in or on the eye. Retinal bleeding and burns may result.
Source facts
Action
On June 3, 2004, the firm had contacted each of its consigness individually via phone, fax, or both.
Initiated
2004-06-03
Root cause
Other
Terminated
2005-01-11
Distribution
The product was distributed directly to health professionals or hospitals.
Recall number
Z-1066-04
Recall status
Terminated
CDRH record ID
33608
FDA classified
2004-07-20
Recall event number
29367