The analysis showed that the incident was caused by a chain of four faults:
1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position.
2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area.
3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation).
4.) Delayed activation of the emergency stop.
Source facts
Action
Consignees were notified of the recall via letter starting on 2/19/16. The letter explains the issue and actions to be taken by the consignee and manufacturer.
Initiated
2016-02-19
Root cause
Device Design
Distribution
AL
AZ
CA
CO
CT
DC
FL
IA
IL
IN
LA
MA
MD
MI
MN
MO
MS
NJ
NY
OH
PA
Puerto Rico
TN
TX
UT
WI
Recall number
Z-1066-2017
Recall status
Open, Classified
CDRH record ID
144238
FDA record created
2017-01-20
Recall event number
73459