Open original FDA record ↗Recall record
- FDA identifier
- Z-1066-2025
- Company in FDA record
- Draeger Medical Systems, Inc.
- FDA classification date
- Jan 30, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.
Source facts
- Action
- Draeger notified consignees on about 01/14/2025 via Urgent - Medical Device Correction letter. Consignees were instructed to inspect the device to identify if the speaker is configured with the ICS in USB mode or Analog mode. For the speaker set to Analog mode there is an analog cable (aux cable) connected to the backside of the speaker and a light on frontside of the speaker display. If the light is green, there is no issue of speaker audio annunciation. If the light is red, the issue of speaker audio annunciation is present. If an issue with alarm annunciation occurs while operating, follow the steps included in the letter. Ensure all affected users or customers, if further distributed, are notified of the recall notification. They were instructed to complete and return the provided reply card.
- Initiated
- 2025-01-14
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AS, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MA, NE, NY, OK, PA, SD, TN, TX, VA, WA and the countries of Argentina, Australia, Austria, Belgium, Botswana, Brazil, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kenya, Malaysia, Mexico, Monaco, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Romania, Samoa, American, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkiye, United Kingdom.
- Recall number
- Z-1066-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211886
- FDA classified
- 2025-01-30
- Recall event number
- 96057