Skip to content

Public FDA record

Z-1067-2015: EPWorks software used in the Protektor 32 Product Usage: Uses electroencephalography (EEG), evoked poten

Recall record Z-1067-2015 for Natus Medical Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1067-2015
Company in FDA record
Natus Medical Incorporated
FDA classification date
Feb 9, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software error occurs when using remote monitoring; if the remote user tries to stop the free run waveform group, the system will display a message informing the user that they do not have sufficient privilege.

Source facts

Action
Natus Medical Inc. DBA Excel-Tech Ltd (Xltek) issued medical device notification letters dated December 22, 2014 to their customers vial email and via courier service on January 2, 2015. Within the notice, customers are instructed are on to removal the previous installation and also on how to install the new program upgrade. Natus Medical Inc. has requested that the Previous Version Removed from Use Confirmation form be completed by each customer and returned. Customers who require assistance can contact Natus Technical service by calling 800-387-7516 or via email Oakville_Technical_Service@xltek.com or ots@natus.com.
Initiated
2015-01-02
Root cause
Software design
Terminated
2017-03-22
Distribution
US Nationwide Distribution
Recall number
Z-1067-2015
Recall status
Terminated
CDRH record ID
132356
FDA classified
2015-02-09
Recall event number
70157

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗