Z-1067-2019: Lithium-ion Rechargeable Batteries for the Philips MX40 Wearable Monitor REF: 989803176201 - Product Usag

Recall record Z-1067-2019 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1067-2019
Company in FDA record
Philips North America, LLC
FDA record creation date
Mar 28, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lithium-ion Rechargeable Batteries for the Philips MX4O Wearable Monitor may have a manufacturing defect that may affect the operation of the monitor

Source facts

Action
Philips issued an email on January 25, 2019 informing customers the batteries on hold had been inadvertently shipped. Philips is asking customers to follow the Action to be Taken by Customer/User should check all MX40s that may contain an affected battery, immediately remove any affected batteries found and replace them with unaffected batteries. The batteries should be returned with the RMA number and return label provided by Philips following any special requirements for shipping lithium-ion batteries specified by the carrier they select
Initiated
2019-01-25
Root cause
Process control
Terminated
2020-04-21
Distribution
US Nationwide distribution in the states of AK, CA, CO, IN,MI, NC, ND, NM, OR, TN, TX and UT.
Recall number
Z-1067-2019
Recall status
Terminated
CDRH record ID
171236
FDA record created
2019-03-28
Recall event number
82264

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗