Open original FDA record ↗Recall record
- FDA identifier
- Z-1067-2019
- Company in FDA record
- Philips North America, LLC
- FDA record creation date
- Mar 28, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Lithium-ion Rechargeable Batteries for the Philips MX4O Wearable Monitor may have a manufacturing defect that may affect the operation of the monitor
Source facts
- Action
- Philips issued an email on January 25, 2019 informing customers the batteries on hold had been inadvertently shipped. Philips is asking customers to follow the Action to be Taken by Customer/User should check all MX40s that may contain an affected battery, immediately remove any affected batteries found and replace them with unaffected batteries. The batteries should be returned with the RMA number and return label provided by Philips following any special requirements for shipping lithium-ion batteries specified by the carrier they select
- Initiated
- 2019-01-25
- Root cause
- Process control
- Terminated
- 2020-04-21
- Distribution
- US Nationwide distribution in the states of AK, CA, CO, IN,MI, NC, ND, NM, OR, TN, TX and UT.
- Recall number
- Z-1067-2019
- Recall status
- Terminated
- CDRH record ID
- 171236
- FDA record created
- 2019-03-28
- Recall event number
- 82264