Z-1068-2014: Renasys EZ PLUS, a Canister Component of of Renasys EZ Negative Pressure Wound Therapy System

Recall record Z-1068-2014 for Smith & Nephew Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1068-2014
Company in FDA record
Smith & Nephew Inc.
FDA classification date
Feb 21, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Modification of the bacterial overflow guard (filter) and related changes to product labeling.

Source facts

Action
The correction strategy is as follows: " An Urgent Medical Device Correction Notice has been prepared to be distributed to all US consignees via Federal Express delivery for confirmation of the delivery. " A copy of the Urgent Medical Device Correction Notice is to be posted on Smith & Nephew's websites, URL corporate, http://www.smith-nephew.com/news-and-media/, and myrenasys.com, concurrent with the distribution of the Notice to consignees. " US distribution of the original design of Canister Kits codes (66800912, 66800913, 66801066, 66800423 and 66800058) subject to this correction ceased November 1, 2013. " The correction has been implemented on a change being affected basis. The modified design is being distributed to US customers effective November 1, 2013. No product will be removed from the market. " Periodic progress reports to FDA, as required per 21 CFR 806 will be submitted monthly until completion of the correction actions..
Initiated
2013-11-14
Root cause
Device Design
Terminated
2014-08-04
Distribution
Worldwide Distribution: USA (nationwide) including states of: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NH, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, VA, WA, WV, and Puerto Rico; and the countries of: Austria, Australia, Belgium, Switzerland, China, Cyprus, Germany, Denmark, Spain, Finland, France, UK, Greece, Israel, Italy, Malta, Netherlands, Norway, Paraguay, Russia, Saudi Arabia, Sweden, Thailand, Turkey, South Africa, Mexico, and Canada.
Recall number
Z-1068-2014
Recall status
Terminated
CDRH record ID
124060
FDA classified
2014-02-21
Recall event number
66893

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗