Open original FDA record ↗Recall record
- FDA identifier
- Z-1068-2024
- Company in FDA record
- OriGen Biomedical, Inc.
- FDA classification date
- Feb 9, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
Source facts
- Action
- On 11/19/23, recall notices were distributed to customers who were asked to do the following: 1) Identify and quarantine affected devices. 2) Further distribution or use should cease immediately. 3) Return affected devices to the recalling firm. 4) Complete and return the customer reply form via email to info@origen.com Customers with questions can contact the recalling firm at 1-800-233-9014 between the hours of 08:00 and 16:30 (CST). On 1/3/24, the firm sent out an additional recall communication removing one of the affected lot numbers.
- Initiated
- 2023-12-19
- Root cause
- Nonconforming Material/Component
- Distribution
- US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE
- Recall number
- Z-1068-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205363
- FDA classified
- 2024-02-09
- Recall event number
- 93752