Skip to content

Public FDA record

Z-1068-2024: Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60m

Recall record Z-1068-2024 for OriGen Biomedical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1068-2024
Company in FDA record
OriGen Biomedical, Inc.
FDA classification date
Feb 9, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.

Source facts

Action
On 11/19/23, recall notices were distributed to customers who were asked to do the following: 1) Identify and quarantine affected devices. 2) Further distribution or use should cease immediately. 3) Return affected devices to the recalling firm. 4) Complete and return the customer reply form via email to info@origen.com Customers with questions can contact the recalling firm at 1-800-233-9014 between the hours of 08:00 and 16:30 (CST). On 1/3/24, the firm sent out an additional recall communication removing one of the affected lot numbers.
Initiated
2023-12-19
Root cause
Nonconforming Material/Component
Distribution
US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE
Recall number
Z-1068-2024
Recall status
Open, Classified
CDRH record ID
205363
FDA classified
2024-02-09
Recall event number
93752

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗