Z-1068-2025: Brand Name: SenTiva DUO" Product Name: VNS Therapy SenTiva DUO", Model 1000-D Model/Catalog Number: 1000-

Recall record Z-1068-2025 for LivaNova USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1068-2025
Company in FDA record
LivaNova USA, Inc.
FDA classification date
Jan 30, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.

Source facts

Action
On January 3, 2025, LivaNova issued a "Urgent Medical Device Removal" via Mail to affected consignees. LivaNova ask consignees to take the following actions: 1. Monitor the patient for changes in clinical symptoms or if the patient loses perception of stimulation (e.g., the typical cadence for patient visits range from every 3-12 months). 2. Perform system diagnostic testing at each patient office visit to monitor generator battery status. 3. Counsel patients to do the following: i. Notify you if there is a change in perceived clinical symptoms ii. Notify you if they no longer perceive any form of stimulation. iii. Patients with epilepsy who have Magnet Mode enabled should be reminded to use the patient magnet daily to check for the sensation of stimulation to confirm proper generator function, in alignment with recommendations in product labeling for all VNS generator models. Notification Acknowledgement Follow one of the options below to acknowledge receipt of this notification: Option 1 1. Using the QR Code or website address and access code on page 1 of this letter, access your customer page via our secure customer portal ( TRIEVR ) to provide the required information to LivaNova USA, Inc. via a secure, transmission-encrypted system. 2. Answer the question in the Response section indicating your receipt and acknowledgement of the notification. 3. Submit the response by selecting the Send Response button at the bottom of the page.
Initiated
2025-01-03
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR,PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT and the countries of AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN,IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA, Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Iceland, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Luxembourg, New Zealand, Norway, Panama, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, The Netherlands, Turkey, UAE, United Kingdom.
Recall number
Z-1068-2025
Recall status
Open, Classified
CDRH record ID
212012
FDA classified
2025-01-30
Recall event number
96063

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗