Open original FDA record ↗Recall record
- FDA identifier
- Z-1069-2025
- Company in FDA record
- Murata Vios, Inc.
- FDA classification date
- Jan 30, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
Source facts
- Action
- Murata sent an email to its consignees on 04/22/2024 to inform them of the issue and provide the correct workflow to address the issue. The email further requested the posting of the workflow guidance in nurse training areas, add to in-services, and other recurrent training for Vios Monitoring. If the flickering issue is observed on the central station monitor, immediately begin using the bedside tablet for all patient monitoring and alarm management and contact Vios Support at 866-990-1434.
- Initiated
- 2024-04-22
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution in the states of NY, TX, OH, MD and the country of India.
- Recall number
- Z-1069-2025
- Recall status
- Open, Classified
- CDRH record ID
- 212211
- FDA classified
- 2025-01-30
- Recall event number
- 96169