Open original FDA record ↗Recall record
- FDA identifier
- Z-1070-2019
- Company in FDA record
- ICU Medical, Inc.
- FDA record creation date
- Mar 28, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.
Source facts
- Action
- On 11/27/18, Urgent Medical Device Recall notices were sent via traceable mailing to customers informing them to discontinue use and distribution of the affected product. In addition, customers were asked to quarantine and return affected product, and to complete and return response forms. Customers were instructed to inform all potential users of the product within their organization, and if the product had been further distributed, to immediately notify these additional customers. Customers with additional questions were encouraged to call: Global Complaint Management @ 1-844-654-7780 (ProductComplaintsPP@icumed.com), Medical Information @ 1-800-241-4002 option 6 (medinfo_us@icumed.onmicrosoft.com), or Customer Care @ 1-949-366-4208.
- Initiated
- 2018-11-27
- Root cause
- Process design
- Terminated
- 2022-02-01
- Distribution
- U.S.: CA, TX, WI, NY, NC, GA, PA, and NY. Foreign: Canada
- Recall number
- Z-1070-2019
- Recall status
- Terminated
- CDRH record ID
- 169674
- FDA record created
- 2019-03-28
- Recall event number
- 81743