Z-1070-2024: VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00

Recall record Z-1070-2024 for Ventec Life Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1070-2024
Company in FDA record
Ventec Life Systems, Inc.
FDA classification date
Feb 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.

Source facts

Action
On 12/22/2023, the firm sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION via Direct overnight mail, or FedEx, or email, or hand delivered, to customers informing them that Ventec Life Systems is recalling a specific lot (listed below) of the VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) since the bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Customer are instructed to: 1. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 2. Stop use of the affected Patient Breathing Package and replace it with a non-affected product. 3. Inspect current stock and quarantine/segregate any unused affected products to prevent their use. For questions/assistance - contact Ventec Life Systems directly at 844-MYVOCSN (844-698-6276), Monday through Friday between 8:00 a.m. - 5:00 p.m. PDT.
Initiated
2024-01-09
Root cause
Process control
Distribution
U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan
Recall number
Z-1070-2024
Recall status
Open, Classified
CDRH record ID
205481
FDA classified
2024-02-23
Recall event number
93806

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗