Z-1073-2022: 1 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h

Recall record Z-1073-2022 for Daavlin Distributing Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1073-2022
Company in FDA record
Daavlin Distributing Company
FDA record creation date
May 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.

Source facts

Action
Beginning April 26, 2022 Daavlin Distributing Co. is sent a recall notification letter via certified mail to customers. Customers are to acknowledge receipt of the notification and a company representative will call customers to arrange for return of the devices to update software. If you have any questions or concerns, contact the company at (800) 322-8546. Monday through Friday, 8:00 AM to 4:30 PM, EST.
Initiated
2022-04-26
Root cause
Software Design Change
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, WV and the countries of Australia, Canada.
Recall number
Z-1073-2022
Recall status
Open, Classified
CDRH record ID
193083
FDA record created
2022-05-12
Recall event number
89990

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗