The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.
Source facts
Action
Lifescan sent an Urgent Medical Device Voluntary Recall letter dated March 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to identify and hold all affected product in inventory and discontinue distributing. The letter states that a Lifescan sales representative will contact customers to arrange for pick up and replacement.
Initiated
2013-03-11
Root cause
Software design
Terminated
2015-08-03
Distribution
USA Nationwide Distribution
Recall number
Z-1074-2013
Recall status
Terminated
CDRH record ID
116649
FDA classified
2013-04-12
Recall event number
64624