Reason stated in the source

New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.

Source facts

Action
On December 27, 2023, the firm notified affected customers via URGENT MEDICAL DEVICE CORRECTION letters. Customers were informed of the IFU update, which included a New Warning: "To reduce the possibility of fibers being drawn into the Impella, customers should avoid exposing the inlet and cannula section of the Impella Heart Pumps to any surfaces or fluid baths where the device can come into contact with loose or floating fibers." The IFU udpate also includes updates to Potential Adverse Events and Warnings to add potential of cardiac or vascular injury. That part of the recall is classified and tracked under RES 93749. Customers should review the IFU updates and send their response form to the recalling firm. Please forward the recall notice to all impacted parties at your facility as well as any facilities to whom subject product has been forwarded. Post a copy of the notice in a visible area for awareness. Customers with questions may contact recallcoordinators@abiomed.com.
Initiated
2023-12-27
Root cause
Labeling Change Control
Distribution
Worldwide distribution - US Nationwide and the countries of AU, CA, DE, FR, IN, MX, TW, and VI.
Recall number
Z-1074-2024
Recall status
Open, Classified
CDRH record ID
205354
FDA classified
2024-02-12
Recall event number
93746

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗