Skip to content

Public FDA record

Z-1075-2018: Iridex TruFocus LIO Premiere" Laser Indirect Ophthalmoscope ( LIO ) The LIO is a headmounted indirect

Recall record Z-1075-2018 for Iridex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1075-2018
Company in FDA record
Iridex Corporation
FDA record creation date
Mar 16, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was discovered that the laser energy was not going where intended to put it, resulting in focal cataracts and iris burns.

Source facts

Action
Action Taken by IRIDEX: 1. A investigation was started immediately upon receipt of the returned devices from the Sloan Kettering complaint on November 3, 2017. CAPA 0125 was initiated to document the events, root cause, correction, corrective and preventive action and effectiveness of these actions. 2. A ship hold was initiated on November 15, 2017. 3.As described below, IRIDEX plans to notify customers and distributors that it is voluntarily recalling all models of the TruFocus LIO Premiere. 4. Public Warning A press release was prepared and sent out on February 23, 2018. For further questions please call (650) 962-8848 Ext. 3672
Initiated
2018-02-27
Root cause
Device Design
Terminated
2023-09-22
Distribution
Worldwide Distribution - USA Distribution and to the countries of : China, Germany, Australia, Canada, Ireland, Lebanon, Japan, Israel and Qatar
Recall number
Z-1075-2018
Recall status
Terminated
CDRH record ID
162191
FDA record created
2018-03-16
Recall event number
79386

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗