Open original FDA record ↗Recall record
- FDA identifier
- Z-1075-2018
- Company in FDA record
- Iridex Corporation
- FDA record creation date
- Mar 16, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was discovered that the laser energy was not going where intended to put it, resulting in focal cataracts and iris burns.
Source facts
- Action
- Action Taken by IRIDEX: 1. A investigation was started immediately upon receipt of the returned devices from the Sloan Kettering complaint on November 3, 2017. CAPA 0125 was initiated to document the events, root cause, correction, corrective and preventive action and effectiveness of these actions. 2. A ship hold was initiated on November 15, 2017. 3.As described below, IRIDEX plans to notify customers and distributors that it is voluntarily recalling all models of the TruFocus LIO Premiere. 4. Public Warning A press release was prepared and sent out on February 23, 2018. For further questions please call (650) 962-8848 Ext. 3672
- Initiated
- 2018-02-27
- Root cause
- Device Design
- Terminated
- 2023-09-22
- Distribution
- Worldwide Distribution - USA Distribution and to the countries of : China, Germany, Australia, Canada, Ireland, Lebanon, Japan, Israel and Qatar
- Recall number
- Z-1075-2018
- Recall status
- Terminated
- CDRH record ID
- 162191
- FDA record created
- 2018-03-16
- Recall event number
- 79386