Z-1075-2021: Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19S, 1/19 GA Flexcore Biopsy Ne

Recall record Z-1075-2021 for Argon Medical Devices, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1075-2021
Company in FDA record
Argon Medical Devices, Inc
FDA record creation date
Feb 5, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.

Source facts

Action
On 12/11/2020, Argon Medical Devices, Inc. distributed an "Urgent-Product Recall Notice" to all affected consignees via US mail. In addition to informing consignees about the recall device, the firm ask consignees to take the following actions: 1. Complete the response form and return it to us as quickly as possible. This will allow us to begin staging replacement product for you and minimize interruption to service. 2. All affected product should be returned to our Argon Athens facility using RGA#______, attention Arbee Cummings. The mailing address is as follows: RGA# ____, Argon Medical Devices, Inc.1445 Flat Creek Road, Athens, TX 75751 USA 3. Argon Medical will ship your replacement devices as soon as we receive your returned product. Your assistance in accounting for the affected devices in your possession is greatly appreciated. If you have any questions about this letter or about the recall action it describes, please contact Beckie.Ellis@argonmedical.com. You may also contact Ms. Arbee Cummings at Arbee.Cummings@argonmedical.com or Ms. Andrea Wieczor at Andrea.Wieczor@argonmedical.com.
Initiated
2020-12-10
Root cause
Process change control
Terminated
2023-01-31
Distribution
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain
Recall number
Z-1075-2021
Recall status
Terminated
CDRH record ID
185023
FDA record created
2021-02-05
Recall event number
87000

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗