Skip to content

Public FDA record

Z-1075-2024: Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.

Recall record Z-1075-2024 for Microbiologics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1075-2024
Company in FDA record
Microbiologics Inc
FDA classification date
Feb 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.

Source facts

Action
The recalling firm issued letters dated 1/8/2024 beginning 1/8/2024 informing the consignee of the issue and providing the following instructions: (1) Review lab procedures to understand how this information affects your usage; (2) Use or discard depending on your lab procedures and how this information affects your usage; (3) Complete the response form provided; (4) Return the response form to the firm; (5) keep the letter for your records; and (6) contact Microbiologics if a replacement kit is needed. On 1/25/2024, the firm issued letters dated 1/24/2024 via email to notify consignees they have expanded the recall to include additional lot numbers. Other than providing additional lot numbers, the letter was the same as the initial letter issued on 1/8/2024. The firm reported the email that had the letter attached alerted the consignee that the recall was expanded to include additional lot numbers.
Initiated
2024-01-08
Root cause
Process control
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NE, NH, NJ, NY, OH, PA, RI, TN, TX, VA, WA, WI, WV, and WY. There was government distribution and no military distribution. Foreign distribution was made to France.
Recall number
Z-1075-2024
Recall status
Open, Classified
CDRH record ID
205457
FDA classified
2024-02-12
Recall event number
93801

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.