Mallinckrodt discovered that some devices may have been manufactured with an out-of-tolerance manometer which can result in in an inaccurate proportional valve calibration table in an affected device.
Source facts
Action
Recall letter was sent to customers in a e-mail beginning November 21, 2016 with a had copy sent via FedEx overnight with signature receipt beginning December 5, 2016. Devices will be picked up and replaced by Mallinckrodt. Questions can be directed to 1-877-KNOW-INO (1-877-566-9466).
Initiated
2016-11-21
Root cause
Equipment maintenance
Terminated
2017-12-20
Distribution
US Distribution to the states of : CA, CO, DC, FL, GA, HI, IA, IL, IN, MI, MN, NC, NJ, NM, NY, OH, PA, and TX
Recall number
Z-1076-2017
Recall status
Terminated
CDRH record ID
151520
FDA record created
2017-01-23
Recall event number
75878