Z-1076-2022: Atrium Advanta V12 Covered Stent System (OUS only). Product Code/Part Number: 85320 85321 85322 85323 85

Recall record Z-1076-2022 for Atrium Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1076-2022
Company in FDA record
Atrium Medical Corporation
FDA classification date
May 18, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible administration of additional anesthesia and contrast which may impact patients' renal function. Potential harms include occlusion or embolism and associated response, with specific outcomes such as amputation, embolism, loss of organ function, organ infarction, or tissue infarction. ***Updated June 2023*** Additional surgical stress caused by prolonged interventional surgery has the potential to lead to myocardial infarction or death.

Source facts

Action
Atrium/Getinge issued Urgent Field Safety Notification to Hospitals (OUS) via on 3/03/22 via email. Letter states reason for recall, health risk and action to take: The company is updating the Advanta V12 Covered Stent System Instructions for Use (IFU). REVISED: Deflation and Withdrawal Instructions: Deflate balloon by pulling vacuum on the inflation device to its maximum volume and allow sufficient time for full deflation. IMPORTANT: Visually verify full deflation of the balloon via fluoroscopy before attempting to withdraw the delivery system. You may continue to use the Advanta V12 Covered Stent Systems, with the consideration of this revised information, which is included on the following page. No Devices Need to Be Returned. If you have any questions, please contact your local Getinge representative. ***UPDATE APRIL 2022*** On April 22, 2022, the firm expanded its recall to include the iCast Covered Stent System. If the device is inflated/deflated using higher concentrations of contrast media or other viscous fluid, as may occur in off-label endovascular procedures, a balloon may take longer to deflate than prescribed in the iCast IFU. Letters were issued to consignees via FedEx with delivery receipt. Customers were informed they may continue to use the iCast covered stent with the consideration of the Deflation and Withdrawal Instructions when used for endovascular procedures. ***UPDATE JUNE 2023*** On June 13, 2023, the firm sent an updated letter to customers to reflect an updated Risk to Health due to new complaint data. ALL LOTS of product, including lots manufactured after May 2020, are affected.
Initiated
2022-03-03
Root cause
Device Design
Distribution
Foreign: Australia Austria Bahrain Belgium Bosnia-Herzegovina Brazil Bulgaria Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt El Salvador Estonia Finland France Germany Great Britain Greece Hong Kong Hungary Iceland Ireland, Republic of Israel Italy Japan Jordan Kuwait Latvia Lebanon Lithuania Macedonia Malaysia Malta Mexico Netherlands New Zealand Norway Panama Paraguay Peru Philippines Poland Portugal Qatar Reunion Russia San Marino Saudi Arabia Singapore Slovak Republic Slovenia South African Republic South Korea Spain Sweden Switzerland Taiwan, Thailand Trinidad and Tobago Turkey United Arab Emirates Uruguay Vietnam
Recall number
Z-1076-2022
Recall status
Open, Classified
CDRH record ID
192681
FDA classified
2022-05-18
Recall event number
89852

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗