Open original FDA record ↗Recall record
- FDA identifier
- Z-1077-2017
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Jan 23, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
under specific settings, an issue may occur during simultaneous Sensor Cartridge and Fluid Pack change on the cobas b 123 <2> POC system and cobas b 123 <4> POC system. The issue occurs when the software function [AutoQC as follow-up] is configured to run all three levels of AutoQC only after a Fluid Pack change, but not after a Sensor Cartridge change. When both are changed simultaneously, starting with the Sensor Cartridge and followed by the Fluid Pack, the analyzer carries out only the follow-up actions associated with the Sensor Cartridge change after completing the change workflow. As a result, no follow-up AutoQC is performed and the three expected AutoQC measurements for the Fluid Pack change are not carried out. Without running quality control, there is a remote possibility that system issues would not be detected and wrong results would not be excluded on all parameters: pH, PO2, PCO2, Na+, K+, Ca++, Cl-, Glu, Lac, Hct, SO2, O2Hb, COHb, MetHb, HHb, and Bili.
Source facts
- Action
- Roche issued an Analyzer Bulletin on September 3, 2015, to all affected Consignees. The recall notifications included a description of the reason for the recall, affected product, and consignee responsibilities; the notification did not include instructions for responding to the notification. Actions Required " Follow the workaround outlined in this Analyzer Bulletin to avoid the described issue. " File this Analyzer Bulletin for future reference. Questions Please contact the Roche Support Network Customer Support Center at 1-800-526-2272 if you have questions about the information contained in this Analyzer Bulletin.
- Initiated
- 2015-09-03
- Root cause
- Software design
- Terminated
- 2017-06-05
- Distribution
- Nationwide Distribution including CA, LA, NE, PR, SC, TX, VA, and WV
- Recall number
- Z-1077-2017
- Recall status
- Terminated
- CDRH record ID
- 151082
- FDA classified
- 2017-01-23
- Recall event number
- 75709