Z-1077-2017: Cobas b 123 POC system The cobas b 123 POC system is a fully automated POC system for whole blood in vitr

Recall record Z-1077-2017 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1077-2017
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA classification date
Jan 23, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

under specific settings, an issue may occur during simultaneous Sensor Cartridge and Fluid Pack change on the cobas b 123 <2> POC system and cobas b 123 <4> POC system. The issue occurs when the software function [AutoQC as follow-up] is configured to run all three levels of AutoQC only after a Fluid Pack change, but not after a Sensor Cartridge change. When both are changed simultaneously, starting with the Sensor Cartridge and followed by the Fluid Pack, the analyzer carries out only the follow-up actions associated with the Sensor Cartridge change after completing the change workflow. As a result, no follow-up AutoQC is performed and the three expected AutoQC measurements for the Fluid Pack change are not carried out. Without running quality control, there is a remote possibility that system issues would not be detected and wrong results would not be excluded on all parameters: pH, PO2, PCO2, Na+, K+, Ca++, Cl-, Glu, Lac, Hct, SO2, O2Hb, COHb, MetHb, HHb, and Bili.

Source facts

Action
Roche issued an Analyzer Bulletin on September 3, 2015, to all affected Consignees. The recall notifications included a description of the reason for the recall, affected product, and consignee responsibilities; the notification did not include instructions for responding to the notification. Actions Required " Follow the workaround outlined in this Analyzer Bulletin to avoid the described issue. " File this Analyzer Bulletin for future reference. Questions Please contact the Roche Support Network Customer Support Center at 1-800-526-2272 if you have questions about the information contained in this Analyzer Bulletin.
Initiated
2015-09-03
Root cause
Software design
Terminated
2017-06-05
Distribution
Nationwide Distribution including CA, LA, NE, PR, SC, TX, VA, and WV
Recall number
Z-1077-2017
Recall status
Terminated
CDRH record ID
151082
FDA classified
2017-01-23
Recall event number
75709

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗