Open original FDA record ↗Recall record
- FDA identifier
- Z-1078-2015
- Company in FDA record
- Teleflex Medical
- FDA classification date
- Feb 10, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Labeling; The affected product packaging is incorrectly labeled as being a cuffed oral ET tube; The correct product labeling is uncuffed nasal ET tube.
Source facts
- Action
- Teleflex Medical sent an " Urgent Medical Device Recall Notification" dated January 9, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They were instructed to immediately discontinue use and quarantine any of the affected products, return the product and to complete the enclosed Recall Acknowledgement Form even if they do not have the affected product in stock. Distributors were instructed to communicate the recall to any of their customers who may have received product included within the scope of the recall. Have the customers return any affected product to them, together with a completed Recall Acknowledgement Form. Teleflex Medical is committed to providing high quality, safe and effective products. We sincerely apologize for any inconvenience this action may cause your operations. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990.
- Initiated
- 2015-01-09
- Root cause
- Labeling mix-ups
- Terminated
- 2016-09-27
- Distribution
- Worldwide Distribution - US Distribution to the states of FL, GA, IL, KY, MA, MI, MO, MS, OH, TX, WA and WV., and to the countries of Canada and the Netherlands.
- Recall number
- Z-1078-2015
- Recall status
- Terminated
- CDRH record ID
- 132631
- FDA classified
- 2015-02-10
- Recall event number
- 70248