Z-1078-2022: Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW.

Recall record Z-1078-2022 for Getinge Usa Sales Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1078-2022
Company in FDA record
Getinge Usa Sales Inc
FDA record creation date
May 13, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and hazardous situations.

Source facts

Action
An Urgent Medical Device Notification was issued 04/01/2022 via FedEx which includes serial numbers for affected devices. Customers are to verify they have an affected device and submit the attached Response Form. Distributors are to forward notification on to customers. Getinge service technicians will contact customers to schedule corrective actions. You can continue to use the Getinge Series 9100E-series Washer Disinfector with G1 control systems until the corrective action is performed, there are no precautionary actions necessary. Contact your Getinge representative with questions or for Technical Support call (888) 943-8872 (options 4,2,1), Monday through Friday, 8:00 a.m. - 6:00 p.m EST, or email techsupportIC.us@getinge.com.
Initiated
2022-04-01
Root cause
Other
Distribution
Worldwide distribution: (US) AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, SD, TX, VA, WA, WV. (OUS) Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Italy, Japan, Kuwait, Libya, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-1078-2022
Recall status
Open, Classified
CDRH record ID
193099
FDA record created
2022-05-13
Recall event number
89998

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗