Z-1079-06: Hitachi Emission Computed Tomography System Hitachi CX-4 CT (including SceptreP3 PET-CT)

Recall record Z-1079-06 for Hitachi Medical Systems America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1079-06
Company in FDA record
Hitachi Medical Systems America Inc
FDA classification date
Jun 3, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The device has a software anomaly which causes blank images to be created during Multiplanar Reconstruction processing or results in an inherent filming error in which a reference scale is improperly sized.

Source facts

Action
Hitachi Medical Systems America, Inc. was informed of a software-related problem with the 50mm scale displayed on filmed images. This problem was reported to the firm by their Japanese manufacturer. In cases where the image has been magnified before filming, the 50mm scale is not correctly sized to the magnified image. The firm issued a Device Correctioin Notice to their customers on 2/21/2006. The notice concerns the referenced software problem which results in the resizing of the images, as well as a second previously discovered problem which results in blank multi-planer reconstructions (MPR''s). According to the Device Correction Notice, a software upgrade has already been installed on all of the units in order to correct the MPRs problem [Ref: RES #35081, Correction Removal Report #1528028-01/31/06-003-C] In response to the resizing issue, a second software upgrade [which the firm states address both issues] will be released on or before 2/27/2006, and this software upgrade will be performed by HMSA''s service personnel during each customer''s next scheduled PM. The Device Notification Letter also instructs the customers to call their Field Service Representatives or the HMSA Applications Helpline if they have any questions about the notice, or if any additional problems are encountered.
Initiated
2006-02-21
Root cause
Other
Terminated
2008-11-13
Distribution
The 16 devices were distributed to 11 states: Texas, Florida, Louisiana, Illinois, Connecticut, Ohio, Massachusetts, Wisconsin, Idaho, California, New Jersey
Recall number
Z-1079-06
Recall status
Terminated
CDRH record ID
45240
FDA classified
2006-06-03
Recall event number
35086

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗