Open original FDA record ↗Recall record
- FDA identifier
- Z-1079-2014
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Feb 24, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Glenosphere to the Baseplate.
Source facts
- Action
- In November and December 2012 Sales staff were provided updated surgical technique and associated video to share with current customers, surgeons, and risk managers. October 17, 2013 Distributors ( via E-mail),and Risk Managers and Surgeons (via traceable courier) were sent written Urgent Recall notifications of these labeling changes. Current customers are identified through an invoice search and distribution history for the affected devices for the past 12 months, since 9/1/12.
- Initiated
- 2013-10-17
- Root cause
- Device Design
- Terminated
- 2014-09-12
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Germany, India, Malaysia, China, Singapore, Korea, Taiwan, and Brazil.
- Recall number
- Z-1079-2014
- Recall status
- Terminated
- CDRH record ID
- 122421
- FDA classified
- 2014-02-24
- Recall event number
- 66430