Z-1079-2026: Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azu

Recall record Z-1079-2026 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1079-2026
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Jan 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.

Source facts

Action
Philips notified consignees on about 12/08/2025 via emailed URGENT Medical Device Correction letter. The letter informed consignees that they have identified nine (9) software issues affecting Azurion R3.0 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorized movement, and/or incorrect image content, and/or loss of data. Consignees were instructed to identify all affected units on hand, ensure all users of the affected units are aware of the issue and notify any customers if the unit has been transferred or distributed. Consignees were also instructed to keep the notification with documentation of the system until the issue is resolved, and complete and return the response form. If any of the nine issues are encountered, consignees were asked to contact their local Philips representative and the letter provided recommended actions. ***Updated May 8, 2026*** Previously, Philips planned to address issues 1-7 in software update 3.1.5, while issues 8 and 9 would be addressed separately. Update 3.1.5 has been canceled. Issues 1-9 will now be addressed through software release 3/1/15. Anticipated release date for software version 3.1.15 is Q3 2026. Local Philips representatives will contact each consignee to schedule visits to install the software updates once available.
Initiated
2025-12-08
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de'Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Kosovo, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Palestine, Philippines, Poland, Portugal, R¿union, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
Recall number
Z-1079-2026
Recall status
Open, Classified
CDRH record ID
217572
FDA classified
2026-01-14
Recall event number
98071

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗