Open original FDA record ↗Recall record
- FDA identifier
- Z-1081-2013
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Apr 10, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When using the V5M transducer, the display of the temperature on the ACUSON SC2000 system can be lower than the actual temperature of the V5M transducer. This occurs only if the system hardware contains revision 4 of the MPI Board. There is a potential to cause esophageal burns in patients.
Source facts
- Action
- Siemens Medical Solutions USA Inc. sent a CUSTOMER SAFETY ADVISORY NOTIFICATION letter dated March 28, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A return response form is included for customers to complete and mail to the recalling establishment. Siemens stated additional follow up action will occur, once material is available, where Siemens field service engineers will visit each unit and replace the affected MPI 4 board. For further questions please call (650) 694-5747.
- Initiated
- 2013-03-28
- Root cause
- Device Design
- Terminated
- 2014-02-14
- Distribution
- Worldwide Distribution - USA and countries of Australia, Austria, Bosnia Herzegovina, Belgium, Finland, France, Germany, India, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway, P.R. China, Portugal, Rep.of Korea, Romania, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan and United Kingdom.
- Recall number
- Z-1081-2013
- Recall status
- Terminated
- CDRH record ID
- 117191
- FDA classified
- 2013-04-10
- Recall event number
- 64855