Open original FDA record ↗Recall record
- FDA identifier
- Z-1081-2015
- Company in FDA record
- Lacrimedics Inc
- FDA classification date
- Feb 10, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf life may not be sustained.
Source facts
- Action
- The firm notified their foreign consignees on 11/26/14. The firm made a decision to conduct a domestic recall on 12/4/14 and will notify US consignees Consignees will be provided with a return label and request any devices that they still have be returned to the firm. Any consignees that have gone out of business will be sent a certified letter to their last known address. CORRECTIVE ACTIONS: The firm has halted sales of the device until the drying method can be improved. Any remaining stock has been placed in quarantine and will be scrapped.
- Initiated
- 2014-11-26
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-03-23
- Distribution
- Worldwide Distribution-US (nationwide) including the states of AR, CA, FL, ID, IL, KS, MA, MI, NC, NJ, NY, OR, PA, and TX and to the countries of Australia, Brazil, Czech Republic, France, Ireland, Mexico, Peru, Switzerland, and United Kingdom.
- Recall number
- Z-1081-2015
- Recall status
- Terminated
- CDRH record ID
- 132171
- FDA classified
- 2015-02-10
- Recall event number
- 70036