Open original FDA record ↗Recall record
- FDA identifier
- Z-1082-2010
- Company in FDA record
- Smiths Medical ASD, Inc.
- FDA classification date
- Mar 11, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential for the power cord's prongs to crack and fail at/or inside the plug. Other medical device manufacturers have reported incidents of sparking, charring, and fires from the affected power cords used with their devices. Smiths Medical has received no reports of incidents regarding the use of the affected Electri-Cord power cords with their devices.
Source facts
- Action
- Consignees were sent a Smiths Medical, Inc. "Urgent Medical Device Recall" letter dated January 27, 2010. The letter was addressed to Risk/Safety Managers, Clinicians, Critical Care Physicians, Pediatricians, Pediatric Intensivists, Nenatologists, and Nursing, Pharmacy, Clinical/Biomedical Engineering, and Anesthesia Professionals. The letter described the problem, products involved and provided "Advice on Action to be Taken by the User". Consignees were requested to complete and return the "Urgent Recall Notice Confirmation Form". To return Power Cords, for questions regarding the recall notice or to report any issues with the affected product, please contact the Smiths Medical Technical Service Department at 1-800-558-2345.
- Initiated
- 2010-01-27
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-12-10
- Distribution
- United States (KY and OH) and Canada.
- Recall number
- Z-1082-2010
- Recall status
- Terminated
- CDRH record ID
- 88564
- FDA classified
- 2010-03-11
- Recall event number
- 54482