Open original FDA record ↗Recall record
- FDA identifier
- Z-1082-2013
- Company in FDA record
- Physio Control, Inc.
- FDA classification date
- Apr 10, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
LIFEPAK CR Plus or LIFEPAK Express Automated External Defibrillators (AEDs) are recalled due to the potential to experience an early internal battery depletion issue.
Source facts
- Action
- The firm start sending the "URGENT MEDICAL DEVICE CORRECTION LIFEPAK CR Plus AND LIFEPAK EXPRESS AEDs" letter, dated February 2013, to their consignees. Consignees are advised to check the readiness of the AEDs and should contact Physio if the OK symbol is NOT visible and either the ATTENTION or WRENCH symbols are present. In the event a device is returned to Physio-Control for assessment, the firm will analyze it to determine if it is affected by the early internal battery depletion issue. If it is verified the device is affected, the consignee will be provided with a permanent replacement device. Customers with questions can visit the firm's website www.physio-control-notices.com/CHARGE-PAK or call Technical Support at 1-800-442-1142.
- Initiated
- 2013-02-08
- Root cause
- Component design/selection
- Terminated
- 2015-11-05
- Distribution
- Worldwide Distribution.
- Recall number
- Z-1082-2013
- Recall status
- Terminated
- CDRH record ID
- 116122
- FDA classified
- 2013-04-10
- Recall event number
- 64427