Convenience Kit contains a PVP Prep Pad, received by a supplier, that does not meet the iodine assay level required to support the 36 month shelf life.
Source facts
Action
On October 16, 2017 a PRODUCT CORRECTION letter was issued to customers notifying them that if they had any kits in their inventory, Centurion will supply labels to affix to the front of each kit instructing the user to discard the prep pad upon opening the kit. Other components of the kit may continue to be used and are NOT within the scope of this recall.
Initiated
2017-10-23
Root cause
Nonconforming Material/Component
Terminated
2018-06-21
Distribution
Nationally
Recall number
Z-1082-2018
Recall status
Terminated
CDRH record ID
162166
FDA record created
2018-03-17
Recall event number
79329