Z-1083-2024: MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007

Recall record Z-1083-2024 for SEDECAL SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1083-2024
Company in FDA record
SEDECAL SA
FDA classification date
Feb 14, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.

Source facts

Action
The distributor in the U.S., Philips, issued letters to consignees dated 05-October-2021 and flagged "IMPORTANT PRODUCT NOTICE" via certified mail on 10/14/2021. The letter explained the problem and under what circumstances it can occur, how to identify whether the consignee has the affected product, and the actions that can be taken to minimize the effect of the problem. The consignee is to: (1) Circulate this notice to all users of the device so that they are aware of the issues; (2) Attach this notice as an addendum to the MobileDiagnost wDR IFU for ease of reference; (3) If unintended motion occurs, press Emergency Stop button to prevent further movement, allow unit to dry and once dry, the system will move as intended, and if the problem persists, contact their local Philips representative and reference the FCO number provided; (4) During cleaning, do not use wet wipes, liquids, and do not spray disinfectants directly on the buttons since this may cause electrical short circuits or metal corrosion, and follow the cleaning instructions provided in the Maintenance, Cleaning and Disposal section of the instruction for use. For actions planned by Philips to correct the problem, the notice says Philips is providing this customer letter containing guidance and alternative process steps to mitigate potential issues. An Important Product Notice Acknowledgement Form was attached for completion and return to Philips no later than 30 days from receipt. The form is to confirm receipt of the Important Product Notice, understanding of the issue, and required actions to be taken.
Initiated
2021-10-14
Root cause
Use error
Distribution
US Nationwide distribution. There was government distribution but no military distribution.
Recall number
Z-1083-2024
Recall status
Open, Classified
CDRH record ID
205051
FDA classified
2024-02-14
Recall event number
93651

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗