Skip to content

Public FDA record

Z-1084-2013: Bracket Buccal Tube U7L DB/UL22 Roth MT HK 0/-14/0 Product Usage: Bracket Buccal Tubes are intended

Recall record Z-1084-2013 for Ormco/Sybronendo. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1084-2013
Company in FDA record
Ormco/Sybronendo
FDA classification date
Apr 10, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ormco Corporation is voluntarily recalling one (1) lot of Bracket Buccal Tubes as a result of a packaging error. More specifically, the Bracket Buccal Tubes product package correctly identifies the product as gingival (0) offset brackets, but actually contains distal (10) offset brackets. Use of the affected product may cause unintended rotation of the molars. Although reversible, a delay in

Source facts

Action
Ormco sent an Urgent Medical Device Recall letter dated March 8, 2013 to all affected customers, via USPS 1st class mail. The letter identified the affected product, problem and actions to be taken. The consignees were instructed to complete the Return Form and return any affected product identified in stock. For questions contact Ormco Customer Care.
Initiated
2013-03-08
Root cause
Packaging
Terminated
2013-09-04
Distribution
Worldwide Distribution - USA (nationwide) including the country of EU.
Recall number
Z-1084-2013
Recall status
Terminated
CDRH record ID
116926
FDA classified
2013-04-10
Recall event number
64749

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗