Z-1084-2015: The MAYFIELD Ultra 360 Base Unit for head positioner, one unit per package. Product Usage: This produc

Recall record Z-1084-2015 for Integra LifeSciences Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1084-2015
Company in FDA record
Integra LifeSciences Corporation
FDA classification date
Feb 10, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Through an investigation of complaints, Integra found that there is the possibility for the locking handle mechanism of the Base Units to fail during use.

Source facts

Action
Integra sent an Urgent - Voluntary Medical Device Recall letter dated December 19, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to take the following actions: 1. Review your MAYFIELD equipment and determine if you have any MAYFIELD Ultra 360 Base Units 2. Once identified we recommend that you discontinue use and remove them from service until they are repaired by an authorized Integra Repair Center 3. Complete the attached Recall Acknowledgement and Repair Form 4. Upon receipt of the Recall Acknowledgement and Repair Form, Customer Service will contact you and provide an RMA number and instructions for returning the productor questions call 609-936-6822. For questions contact Customer Service: USA 800-654-2873 Outside USA 609-936-5400
Initiated
2014-12-19
Root cause
Nonconforming Material/Component
Terminated
2017-02-22
Distribution
Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, and DC. (25 government consignees) and in the foreign countries: CANADA, ALGERIA, ARAB EMIRATES, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CHILE, CHINA, COLUMBIA, CYPRESS, DENMARK, DOMINICAN REPUBLIC, FRANCE, GERMANY, GREAT BRITAN, HONG KONG, HUNGARY, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN , JORDAN, KOREA, LIBYA, MACEDONIA, MAYLASIA, MEXICO, NETHERLANDS, POLAND, PORTUGAL, PUERTO RICO, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI LANKA, TAIWAN, THAILAND, and TURKEY.
Recall number
Z-1084-2015
Recall status
Terminated
CDRH record ID
132230
FDA classified
2015-02-10
Recall event number
70088

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗