Z-1086-06: St. Jude AGILIS Steerable Introducer, Reorder number 408304 and 408308 (Australia). Sterile EO. St. Jude

Recall record Z-1086-06 for St. Jude Medical / Daig Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1086-06
Company in FDA record
St. Jude Medical / Daig Division
FDA classification date
Jun 3, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

St Jude Medical has determined that a limited number of Agilis Steerable Introducers manufactured in late 2005 are susceptible to the hemostasis hub leaking or separating from the handle. The risks that exist are related to the loss of hemostasis through the Agilis introducer and/or the possibility of air or foreign body being introduced into the patient with resultant embolism.

Source facts

Action
A Field Action Notification, dated 04/13/06, was sent to hospital's Electrophysiology labs (cath labs) and to physicians. This Notification describes the issue, the risk to patient, identifies lot numbers of product affected and states that a SJM Rep will be contacting them soon. The SJM Reps visited the hospitals to verify they received the notification and retrieve affected product. Replacement product will be provided at no charge. EXPANDED RECALL A Field Action Notification, dated 06/01/06, was sent to hospital's Electrophysiology labs (cath labs) and to the physicians who reported a problem with the device. This Notification describes the issue, risk to patients, identifies lot numbers and states that a SJM Representative will be contacting them soon to facilitate the return of product. Replacement product will be offered at no charge. The physician notification, dated 06/01/2006, describes the issue, the risk, lot number affected and informs them that a notification has been sent to the hospitals EP Lab.
Initiated
2006-04-14
Root cause
Other
Terminated
2007-04-01
Distribution
World wide- USA states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV. ---Foreign includes: Austria, Belgium, Denmark, France, Germany, Great Britain, Italy, Lichtenstein, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland, United Kingdom
Recall number
Z-1086-06
Recall status
Terminated
CDRH record ID
45749
FDA classified
2006-06-03
Recall event number
35300

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗