Z-1086-2024: The Sofia 2 Flu + SARS Antigen FIA employs immunofluorescence technology in a sandwich design that is use

Recall record Z-1086-2024 for Quidel Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1086-2024
Company in FDA record
Quidel Corporation
FDA classification date
Feb 14, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

One lot of the Sofia 2 Flu + SARS Antigen FIA (Lot 709083) was labelled with incorrect expiration dates. The kits contain pouched cassettes labeled with an expiration date of May 31, 2025, however the correct expiration date for the pouches is September 12, 2024. The kits are labeled with an expiration date of April 30, 2025, however, the correct expiration date for the kits is September 12, 2024. The issue was discovered on November 28, 2023 by a supply chain employee who determined that the kits had expiration dates beyond the FDA authorized shelf life.

Source facts

Action
On 1/3/24, correction notices were emailed to customers and distributors who were asked to do the following: 1) Complete and return the attached Confirmation of Notification form to via fax at 858.203.9297 or email at customernotifications@quidelortho.com 2) Indicate the amount of product to be replaced on the Confirmation of Notification form. You are not required to take any action related to already used kit material and can continue to use this lot number until replacement product arrives. 3) Upon receipt of the replacement product, discontinue using, render unusable and discard the remaining inventory of the affected lot. Contact Technical Support at customernotifications@quidelortho.com or by calling 800.874.1517 (in the U.S.) with any questions or concerns that you may have regarding this issue. In addition distributors were asked to do the following: Provide your list of affected customers that received this lot number. Please include a contact name, email, and phone number for each consignee. Once this list is received, the recalling firm will notify all customers on your behalf, collate responses, and facilitate replacements directly.
Initiated
2024-01-03
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of FL, TN, IL, LA, NY, IN.
Recall number
Z-1086-2024
Recall status
Open, Classified
CDRH record ID
205454
FDA classified
2024-02-14
Recall event number
93799

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗