Reason stated in the source
Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.
Source facts
- Action
- Elekta sent an Urgent Important Field Safety Notice on January 7, 2015, to all affected customers. Elekta recommended that customers replace the M20 x 50 MM bolts marked 'OF' on the gantry counterbalance frame immediately. The digital linear accelerator must not be used until these 'OF' bolts are replaced. Elekta recommended that customers put the machine in a parked position at 0, and attach the gantry locking straps. A Modification kit will be available as soon as possible which contains the correct bolts. Customers were instructed to complete the Important Field Safety Notice Acknowledgement form and return to their local Elekta Office or Representative. Customers with questions were instructed to contact their local Elekta representative. For questions regarding this recall call 404-993-5886.
- Initiated
- 2015-01-07
- Root cause
- Other
- Terminated
- 2016-04-14
- Distribution
- Worldwide Distribution - USA including AZ, CA, FL, KS, LA, MN, MO, NY, NC and Internationally to Austria, Bulgaria, Canada, China, Germany, Hungary, Italy, Malaysia, Malta, Netherlands, Norway, Russia, Spain, Sweden, Taiwan and Ukraine.
- Recall number
- Z-1088-2015
- Recall status
- Terminated
- CDRH record ID
- 132771
- FDA classified
- 2015-02-10
- Recall event number
- 70279