Skip to content

Public FDA record

Z-1088-2024: CombiDiagnost R90 1.1 (709031)

Recall record Z-1088-2024 for PHILIPS MEDICAL SYSTEMS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1088-2024
Company in FDA record
PHILIPS MEDICAL SYSTEMS
FDA classification date
Feb 14, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.

Source facts

Action
An Important Product Notice (2021-PD-DXR-034) dated February 15, 2022 was issued via Certified Mail. Philips Field Service Engineers will contact customers to schedule an onsite visit to install update software correcting the issue. Customers are asked to follow the provided instructions to ensure there are no pending images before an exam circulate the notice to all users and to complete and return the response form and circulate the notice to all users of the device for awareness. Contact Philips Customer Care Solutions Center 1-800-722-9377 with questions.
Initiated
2022-02-11
Root cause
Software design
Distribution
US Nationwide distribution in the states of CA, FL, GA, IL, IN, KS, MD, ME, MI, MN, NC, NJ, NY, TX, VA, WA.
Recall number
Z-1088-2024
Recall status
Open, Classified
CDRH record ID
205554
FDA classified
2024-02-14
Recall event number
93819

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗