Skip to content

Public FDA record

Z-1088-2026: Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF

Recall record Z-1088-2026 for Medline Industries, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1088-2026
Company in FDA record
Medline Industries, LP
FDA classification date
Jan 14, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Source facts

Action
Medline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/16/2025 via email and USPS first class mail. The notice explained the issue, potential risk, and requested the following: "1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantityof affected product you have in inventory on the form. Even if you do not have any affected productin inventory, please complete and submit the form. Website link: https://recalls.medline.comRecall Reference #: R-25-145-FGX5 Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4.If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com."
Initiated
2025-12-16
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-1088-2026
Recall status
Open, Classified
CDRH record ID
217878
FDA classified
2026-01-14
Recall event number
98197

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.