Open original FDA record ↗Recall record
- FDA identifier
- Z-1089-2014
- Company in FDA record
- Orthofix, Inc
- FDA classification date
- Feb 25, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a possibility that the locking mechanism of the Azure Anterior Cervical Plate may fracture during or after implantation.
Source facts
- Action
- Orthofix sent a Urgent Medical Device Recall Notification dated October 29, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Immediate Action: 1. Immediately cease any further distribution or use of the AZURE" plates. 2. Notify your surgeons and staff customers that Azure" plates cannot be used and are to be returned to Orthofix. at no cost to you, see Instructions below. 3. Please complete the attached reply form indicating that you have received this letter. Product Return Instructions: To return the Azure plates, please follow these Instructions. 1. Check you inventory of Azura Implant Trays.. 2. Complete the attached form with the quantity you are returning. 3. Contact your local Orthofix representative or Orthofix customer service representative at 888--298-5700 for return instructions and shipment information and replacement with a different Orthofix Anterior Cervical Plate System. .
- Initiated
- 2013-10-29
- Root cause
- Component design/selection
- Terminated
- 2014-12-09
- Distribution
- Worldwide Distribution: USA (nationwide) and the country of Germany.
- Recall number
- Z-1089-2014
- Recall status
- Terminated
- CDRH record ID
- 122914
- FDA classified
- 2014-02-25
- Recall event number
- 66660