Skip to content

Public FDA record

Z-1089-2017: Optilite IgG4 Kit Product Code: LK009.OPT.A

Recall record Z-1089-2017 for The Binding Site Group, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1089-2017
Company in FDA record
The Binding Site Group, Ltd.
FDA record creation date
Jul 27, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

A change to the lower end of the measuring range and product insert did not reflect the change.The measuring range provided in instrument parameters is correct but contradicts the information provided in the product insert

Source facts

Action
Customer notification letters were sent on 12/23/16. The Binding Site recommends to continue testing on the kit as normal and refer to the measuring ranges and limit of quantification stated in the notice and as indicated in the amended insert (INS009.OPT.A, Version: 21st December 2016) provided with this notification. The notice needs to be passed on to all those who need to be aware within the organization or to any Organization where the potentially affected devices have been transferred. Return the completed E-Back Form to your local Binding Site Representative for records and note that The Binding Site intends to report this correction to the FDA recall coordinator. If you have any questions contact Technical Services.
Initiated
2016-12-23
Root cause
Under Investigation by firm
Terminated
2017-05-03
Distribution
U.S. distribution to the following; FL, TX, PA, CA, MA, WA, MI, CT. No foreign distribution.
Recall number
Z-1089-2017
Recall status
Terminated
CDRH record ID
152264
FDA record created
2017-07-27
Recall event number
76131

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗