Skip to content

Public FDA record

Z-1090-2017: cobas p 612 pre-analytical system; Pre-analytical sample handling that includes de-capping, Aliquotting

Recall record Z-1090-2017 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1090-2017
Company in FDA record
Roche Diagnostics Corporation
FDA classification date
Jan 24, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Source facts

Action
On 12/27/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected Consignees via courier fax or email. The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The cobas p 612 pre-analytical system (63x) software is being revised to correct the aspiration process. We anticipate this new version (2.0.4) to be available during Quarter 1 2017. For now, discontinue using disposable non-filter tips for Aliquotting on your cobas p 612 pre-analytical system (63x). You will be contacted by a Roche Field Engineering Specialist (FES) about modifying your system to allow for the use of disposable filter tips. Filter tips will be provided at the time of service. Actions Required " Immediately discontinue using disposable non-filter tips for Aliquotting on your cobas p 612 pre-analytical system (63x). " Complete the enclosed faxback form and fax it to number 1-877-906-8982 or email it to Indianapolis.cd_tsc@roche.com. " Provide copies of this UMDC to other clinicians who may need to be aware of this software issue. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336, 24 hours a day/7 days a week, if you have questions about the information contained in this UMDC.
Initiated
2016-12-27
Root cause
Software design
Terminated
2017-08-04
Distribution
TX and WA
Recall number
Z-1090-2017
Recall status
Terminated
CDRH record ID
152104
FDA classified
2017-01-24
Recall event number
76078

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.