Open original FDA record ↗Recall record
- FDA identifier
- Z-1090-2019
- Company in FDA record
- Integra Limited
- FDA record creation date
- Apr 5, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Temperature indicator may have changed color to dark gray or black prior to shipment. Change in color may indicate the product exposure to temperatures greater than 110F
Source facts
- Action
- Integra LifeSciences Corp. notified consignees (Hospitals and Distributors) by letter delivered by courier service, facsimile or email about the initial notification on March 1st, 2019.Letter explains problem, health risk and action to take: examine inventory to determine if you have any affected lots with discolored temperature dots, where the dots look similar as pictures labeled UNACCEPTABLE on previous page.. If you do have any product with discolored temperature dots, stop using the product immediately, remove them from service and place in quarantine until the product can be returned. Complete the attached Acknowledgement and Return Form. Questions Customer Service at 1-800-654-2873 Monday to Friday 8:00 AM 8:00 PM EST or custsvcnj@integralife.com .
- Initiated
- 2019-03-01
- Root cause
- Process control
- Terminated
- 2020-06-22
- Distribution
- US Nationwide distribution in the states of OH, MI.
- Recall number
- Z-1090-2019
- Recall status
- Terminated
- CDRH record ID
- 171443
- FDA record created
- 2019-04-05
- Recall event number
- 82361