Skip to content

Public FDA record

Z-1091-2019: Integra bioBLOCK Resorbable subtaler Implant, 12mm Catalog Number: 040012 - Product Usage: The bioBLOCK

Recall record Z-1091-2019 for Integra Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1091-2019
Company in FDA record
Integra Limited
FDA record creation date
Apr 5, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Temperature indicator may have changed color to dark gray or black prior to shipment. Change in color may indicate the product exposure to temperatures greater than 110F

Source facts

Action
Integra LifeSciences Corp. notified consignees (Hospitals and Distributors) by letter delivered by courier service, facsimile or email about the initial notification on March 1st, 2019.Letter explains problem, health risk and action to take: examine inventory to determine if you have any affected lots with discolored temperature dots, where the dots look similar as pictures labeled UNACCEPTABLE on previous page.. If you do have any product with discolored temperature dots, stop using the product immediately, remove them from service and place in quarantine until the product can be returned. Complete the attached Acknowledgement and Return Form. Questions Customer Service at 1-800-654-2873 Monday to Friday 8:00 AM 8:00 PM EST or custsvcnj@integralife.com .
Initiated
2019-03-01
Root cause
Process control
Terminated
2020-06-22
Distribution
US Nationwide distribution in the states of OH, MI.
Recall number
Z-1091-2019
Recall status
Terminated
CDRH record ID
171444
FDA record created
2019-04-05
Recall event number
82361

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗