Skip to content

Public FDA record

Z-1091-2022: ATLAS Stim Headbox-NK For use only with the ATLAS Neurophysiology System. CAUTION--INVESTIGATIONAL DEVICE

Recall record Z-1091-2022 for Neuralynx Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1091-2022
Company in FDA record
Neuralynx Inc
FDA record creation date
May 13, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Atlas Stim Headbox (ASHB) was delivered to US customers under a premise that the device had an Investigational Device Exemption (IDE) and was for research use only, but the manufacturer does not have an approved IDE by the FDA. The product has not undergone electrical safety testing and has not received regulatory clearance.

Source facts

Action
Customer notifications were sent out to all affected customers on March 14, 2022, via USPS certified mail. Actions to be taken: Field representatives will be contacting consignees to return the effected product and coordinate a replacement with their cleared headbox, product models 31-0601-0007, 31-0601-0010 or 31-0601-0017, by April 6, 2022. In the interim, it is advised to stop use of this product. Any further questions regarding this matter, contact by email at kmoeykens@fh-co.com or by telephone at 207-666-5425 Monday -Friday 8:00 AM -4:30 PM EST.
Initiated
2022-03-14
Root cause
No Marketing Application
Distribution
U.S. Nationwide distribution in the states of AL, CA, IL, MN, NC, RI.
Recall number
Z-1091-2022
Recall status
Open, Classified
CDRH record ID
192671
FDA record created
2022-05-13
Recall event number
89843

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗