Open original FDA record ↗Recall record
- FDA identifier
- Z-1092-2014
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Feb 26, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal. The other two zones exceeded the minimum specification.
Source facts
- Action
- Zimmer sent an Urgent Medical Device Recall dated January 10, 2014 to distributors via electronic mail. The hospital/surgeon that has implanted this manufacturing lot will be identified and notified by the distributor. The letter identified the affected product, problem and actions to be taken. Distributors were instructed to locate and remove the affected product in their territory, as well as identifying and notifying the hospital/surgeon that has implanted this manufacturing lot as applicable. The distributor will return on-hand affected products to Zimmer and ensure all of their products are accounted for using the return form provided in the letter. Distributors and hospital consignees are to complete and return the inventory return form and acknowledgement form indicating which hospital/surgeon has implanted the product and to which they have provided notice to. Email a copy of the completed return form to CorporateQuality.PostMarket@Zimmer.comCommunications outside of the United States will occur approximately two weeks after the United States communications. For questions call 1-877-946-2761
- Initiated
- 2014-01-10
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-09-11
- Distribution
- Distribution USA in the state of Illinois.
- Recall number
- Z-1092-2014
- Recall status
- Terminated
- CDRH record ID
- 124650
- FDA classified
- 2014-02-26
- Recall event number
- 67067