Z-1092-2022: Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system

Recall record Z-1092-2022 for Luminex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1092-2022
Company in FDA record
Luminex Corporation
FDA classification date
May 13, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.

Source facts

Action
On 04/21/22 Luminex issued a "Urgent Medical Device Recall" Notification to affected consignees. On 6/16/22 and 7/14/22 Luminex expanded the recall to include additional serial numbers. Luminex is asking consignees to take the following actions: 1. Continue use of your VERIGENE Processor SP as service has been completed by a Luminex Field Service Engineer and the Processor SP s heaters are now within specification. 2. Perform a retrospective review of results for VERIGENE Gram-Positive Blood Culture (BC-GP) and VERIGENE Gram-Negative Blood Culture (BC-GN) Nucleic Acid Tests from 18 months prior to the most recent service visit. Identify any results that were not consistent with other clinical signs and symptoms. Promptly inform Luminex of any previously unreported incidences of false results. 3. Distribute the notification to any individuals within your organization who need to be made aware. 4. Please reach out to Global Support Services if a copy of the service record is required. 5. Please complete the Acknowledgement & Receipt form and return within 10 business days.
Initiated
2022-04-05
Root cause
Under Investigation by firm
Distribution
US:OH, ME, PA, FL, IL, NE, IN, TX, GA, WA, MI, CO, AZ, OH, KY, CA, VT OUS: None
Recall number
Z-1092-2022
Recall status
Open, Classified
CDRH record ID
193171
FDA classified
2022-05-13
Recall event number
90023

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗